Acticare IC - Technical Information |
|
Feature |
Specification |
| Modes: |
IC, TSE, and TENS therapy options
with advanced user mode. |
| IC channels: |
Two 2-pole coupled channels. |
| User controls: |
Full digital control of all
treatment and waveform parameters. |
| IC waveform: |
Pure sine-wave. |
| IC carrier frequencies: |
2kHz and 4kHz |
| IC modulation frequency (AMF)
range: |
0Hz to 250Hz |
| IC sweep patterns: |
6^6, 1/1, 6/6 |
| IC treatment programmes: |
8 preset programs + fully user
programmable mode. |
| Treatment timer: |
User adjustable from 1 - 30
minutes (IC) or 10 - 180 minutes (TSE and TENS). |
| Pulse rate: |
1Hz - 500kHz |
| TSE and TENS pulse form: |
Biphasic, mono-phasic or twin
peak. |
| TSE and TENS pulse duration: |
0.5 to 200 us |
| Patient in circuit detection: |
Limits output and displays
warning message if patient disconnected (TSE and TENS modes only). |
| Power source: |
Four rechargable AA batteries (average
two hour operating life) with mains re-charging cradle
(charge time from flat = 2 hours). |
| Maximum current: |
100mA peak (IC), 20mA average (TSE and TENS) |
| Maximum charge per pulse: |
25uC limited by safety circuit
(all modes) |
| Control: |
Microprocessor controlled with
hardware supervisory safety circuit. |
| Dimensions: |
100mm high, 72mm deep, 38mm
wide |
| Key design compliance: |
IEC 60601-2-10 |
|
|